Mass spectrometry is a method for determining the molecular ions and fragment ions of a sample in a high vacuum system to determine the relative molecular mass and molecular structure of the sample.

The AMI Master 400 Gas Spectrometer provides temperature control to effectively prevent condensation and has the ability to establish automatic control with external instruments. It also features  built-in filament pressure protection to prolong filament life and a Faraday electron multiplier with dual high sensitivity detectors from 100% to ppb.

Purpose designed analytical software together with multi-component sampling for qualitative and quantitative analysis, a customisable sampling system for pre-treatment and multi-channel sampling detection and millisecond response and scanning speed ensure the AMI Master 400 delivers fast online analysis.

Field Applications

  • Thermoanalysis
  • Programmed temperature desorption catalytic reaction
  • Automobile exhaust analysis 
  • Reaction process monitoring
  • Battery production
  • Fermentation
  • Hydrogen purity analysis
  • On-line detection of fuel cell exhaust gas
  • CVD (Chemical Vapor Deposition)

MASTER 400 Enquiry

Specifications

Mass Ranges (AMU)
1-100amu/1-200amu/1-300amu

Mass Analyzer
quadrupole

Detector
Faraday (sem optional)

Detection limit: Faraday
4 X10 -13 mbar

Detection limit: SEM
3 X10 -15 mbar2

Source of Ion
Open ion source with 2 filaments

Measurement channel
128 pieces

Communication Interface
Ethernet

Scan Time
1ms-16s/amu

10% Peak Resolution
0.5-2.5amu

Sample Gas Pressure
Atmospheric pressure (other pressures are optional)