MASTER 320

The ALTAMIRA MASTER 320 is a benchtop gas analysis system capable of sampling gases at or below atmospheric pressures. The standard version is designed for manually controlled as well as automatic analysis of non-corrosive non-flammable gases.  A corrosive version, the MASTER 320C is also available and is suitable for some corrosive and/or flammable gas applications.

Overview

The ALTAMIRA MASTER 320 is comprised of an inlet system, residual gas analyzer, diaphragm vacuum pump and turbo pump.

The inlet system utilises a heated quartz or stainless steel capillary up to 350ºC and aids to prevent of condensation of sample gas vapours during analysis.  The MASTER 320 is capable of sampling in mass ranges of 1-100 amu, 1-200 amu and 1-300 amu.

Features

  • A compact all in one gas analysis system 
  • Intuitive, easy to use display
  • Very low detection limit (<1ppm)
  • Heated capillary inlet (up to 350ºC) to prevent condensation during analysis
  • Mass ranges from 1-100 amu, 1-200 amu, and 1-300 amu
  • Diverse application base with pressure ranges from vacuum to atmosphere
  • Easy bakeout options to remove residual contaminants and reduce background noise

MASTER 320 Enquiry

Specifications

Mass ranges (AMU)
1-100 / 1-200 / 1-300

Gas Connection
Stainless steel capillary / quartz capillary

Capillary diameter (mm)
Stainless steel: 1.59 OD, 0.125 ID
Quartz: 0.22 OD, 0.15 ID

Gas inlet
Stainless steel, FPM

Gas flow rate (sccm)
1-2

Sample Gas Pressure (mbar)
500-1500

Capillary Operating Temperature
200oC (standard)
350oC (maximum)

Storage/Transport
5-45oC

Operation
12-35oC

Altitude
<50 dB in full operation

Detector
Faraday/C-SEM

Scan Speed
MID: 2ms/amu – 60s/amu
Bar graph stair scan: 2ms/amu – 60 s/amu
Analog scan: 20ms/amu – 60 s/amu
Bar graph scan: 20 ms/amu – 60 s/amu

Weight
35 kg

Size (L x W x H)
545 x 305 x 395 mm

Voltage
100, 120, 230 VAC
4 Analog outputs: 4x 0-10 V / 12 Bit
4 Digital inputs: 4x external protection
7 Digital outputs: 7x Sink, optical isolated 24V

Pump Status
Digital Output channel 1, 1 relay, switching contact, NC, NO, COM

Connectors
15-pin Dsub (User I/O) and 25 pin Dsub (Aux I/O)